Skip to main content
LibertyCES — Engineering Line Direct: Call James Riggins at (559) 395-5500 — 30+ Years Spec Experience. No sales queue. Real engineering answers.
Chemical Compatibility

pH/ORP Instrument — Salts/Brines

Moderate evidence

30+ years spec experience. Send James the process data and get a real answer — not a catalog page.

Request a Spec Review →

Named chemicals

Cyanide (CN-)Hexavalent Chromium (Cr6+)

What LibertyCES's vendor documentation shows

PHED explicitly named for electroplating media contaminated with Cr6+ and CN-.

Materials documented for this service

  • Open-ring diaphragm (PHEX/RHEX)

Products cited for this service

ProMinent PHED/PHEX/RHEX

Real Specifications

ProMinent PHED applicationProcess water/electroplating contaminated with Cr6+, CN-prominent-sensors-controllers.md
PHED max temperature/pressure80°C / 8 barprominent-sensors-controllers.md
PHED diaphragm2x ceramic, double junctionprominent-sensors-controllers.md
PHEX/RHEX max temperature/pressure25°C/16bar or 100°C/6barprominent-sensors-controllers.md
PHEX diaphragm designOpen ring diaphragm (high-solids, no chemical contamination service)prominent-sensors-controllers.md
Electroplating rinse water contaminated with cyanide or hexavalent chromium needs a double-junction sensor, not a standard single-junction pH probe — ProMinent's PHED is built specifically with a 2x ceramic double-junction diaphragm for this contaminated-media service, rated to 80°C/8 bar, distinct from the plain PHEP/PHES sensors quoted for clean drinking or process water.

Every material and product claim on this page traces to real, vendor-published documentation on file with LibertyCES. Chemical compatibility depends on exact concentration, temperature, and duty cycle — confirm current specifications with the manufacturer and request a spec review before final specification.

FAQ

Is pH/ORP Instrument compatible with Salts/Brines?

PHED explicitly named for electroplating media contaminated with Cr6+ and CN-. This is moderate evidence.

What materials are documented for pH/ORP Instrument in Salts/Brines service?

Materials documented for this service: Open-ring diaphragm (PHEX/RHEX). Chemical compatibility depends on exact concentration, temperature, and duty cycle — confirm with a spec review before final specification.

What specific Salts/Brines chemicals does this cover?

Named chemicals in this documentation: Cyanide (CN-), Hexavalent Chromium (Cr6+).

What products are cited for pH/ORP Instrument in Salts/Brines service?

Documented examples include ProMinent PHED/PHEX/RHEX.

Ready to buy or need a fast answer? Call, text, or email James directly — (559) 395-5500 · [email protected].